H-335 – Xylem H-335 User Manual
Page 19
H-335
SDI-12 Command and Response Protocol 2-3
Example of a H-335 "aM1!" command:
Command
Response
Time Values
Description
"aM1!"
"a0036
3 sec 6
Make measurement
Subsequent Command
Response
"aD0"
a+AA.AA+BB.BBB+CCCC+DDDD+EEEE+FF.FF
Where:
+AA.AA
= User Angle
+BB.BBB
= Sensor Angle
+CCCC
= Raw A/D Positive Count
(±8192)
+DDDD
= Raw A/D Negative Count (±8192)
+EEEE
= Raw A/D Count
(±16384)
+FF.FF
= Sensor Temperature
(C)
Example of a H-335 "aM2!" command:
Command
Response
Time Values
Description
"aM2!"
"a0031
3 sec 1
Make measurement
Subsequent Command
Response
"aD0"
a+AA.AA
Where:
+AA.AA
= Gate Height
“aM1" Command
“aM2" Command
See also other documents in the category Xylem Equipment:
- H-2191 (57 pages)
- H-223 (9 pages)
- H-264 (16 pages)
- H-310 (40 pages)
- H-312 (38 pages)
- H-3301 (24 pages)
- H-3342 (30 pages)
- H-339 (21 pages)
- H-340-CalKit (6 pages)
- H-340 (12 pages)
- H-340SDI (32 pages)
- H-3521 (57 pages)
- H-3531 FlashLite (76 pages)
- H-3551 (42 pages)
- H-3553 (36 pages)
- H-375 04101 WIND MONITOR-JR (18 pages)
- H-4161 (36 pages)
- H-4191 (21 pages)
- H-424MS V1.0 (25 pages)
- H-424MS V1.1 (22 pages)
- H-4271 (22 pages)
- H-4280 (34 pages)
- H-4400 (7 pages)
- H-4401 (6 pages)
- H-5223 (107 pages)
- STORM 3 UPGRADE (4 pages)
- System 5000 BASIC Manual (54 pages)
- System 5000 User Manual (136 pages)
- DH-21 (112 pages)
- BrazedPlate Heat Exchangers (4 pages)
- Heat Exchangers (8 pages)
- GL500 MULTICHANNEL DATA LOGGER (32 pages)
- 4046 ANALOG TO SDI-12 CONVERTER (7 pages)
- 4015 SDI-12 TO ANALOG CONVERTER (5 pages)
- iRIS-CAM Wastewater Security Camera (11 pages)
- SDI-12 Data Logger iRIS 220, iRIS 320, iRIS 350, iRIS 150, iRIS 300 (111 pages)
- iRIS 150 GPRS DATALOGGERS (53 pages)
- iCE3 GPRS Variant (29 pages)
- GL500-2-1 Data Logger (32 pages)
- 3101 CONDUCTIVITY CONTROLLER (11 pages)
- 3675 pH CONTROLLER (9 pages)
- 378 TEMPERATURE CONTROLLER (9 pages)
- Expert 1400 (16 pages)
- Expert 700 (8 pages)