H-3342 – Xylem H-3342 User Manual
Page 20

3-8 SDI-12 Command and Response Protocol
H-3342
3.7 Send Identification Command
The Send Identification Command responds with sensor vendor, model, and version data. Any
measurement data in the sensor's buffer is not disturbed.
Command
Response
"aI!"
"allccccccccmmmmmmvvvxx...xx
Where:
a
is the sensor address ("0-9", "A-Z", "a-z", "*", "?").
I
is an upper-case ASCII character.
ll
is the SDI-12 version compatibility level, e.g. version 1.2 is represented as
"12".
cccccccc
is an 8 character vendor identification to be specified by the vendor and
usually in the form of a company name or its abbreviation.
mmmmmm
is a 6 character field specifying the sensor model number.
vvv
is a 3 character field specifying the sensor version number.
xx...xx
is an optional field of up to a maximum of 13 characters to be used for
serial number or other specific sensor information not relevant to operation
of the data recorder.
Example of a H-3342 "aI!" command:
"a12 DAA H-334vvvS#nnnnnnVkkk
"
H-3342 implementation of the optional 13 character field:
S#nnnnnnVkkk
(12 bytes total)
Where:
"nnnnnn"
is a six character sensor serial number
"kkk" is a three digit sensor firmware revision level