beautypg.com

H-3342 – Xylem H-3342 User Manual

Page 19

background image

H-3342

SDI-12 Command and Response Protocol 3-7

3.6 Initiate Verify Command
The Verify Command causes a verify sequence to be performed. The result of this command is
similar to the "aM!" command except that the values generated are fixed test data and the results
of diagnostic checksum tests. The data generated in response to this command is placed in the
sensor's buffer for subsequent collection using "D" commands. The data will be retained in the
sensor until another "M", "C", or "V" command is executed.

Command

Response

Description

"aV!"

"atttn"

Initiate verify sequence

Where:

a

is the sensor address ("0-9", "A-Z", "a-z", "*", "?").

V

is an upper-case ASCII character.

ttt

is a three digit integer (000-999) specifying the maximum time, in seconds, the
sensor will take to complete the command and have data available in its buffer.

N

is a single digit integer (0-9) specifying the number of values that will be placed in
the data buffer. If "n" is zero (0), no data will be available using subsequent "D"
commands

Example of a H-3342 "aV!" command:

Command

Response

Time

Values

Description

"aV!"

"a0013"

1 sec

3

Return fixed data and
diagnostic data for testing purposes.

Subsequent Command

Response

"aD0"

a+123.456+78.9+y

Key

Description

Units

+123.456

Fixed test data

+78.9

Fixed test data

y

ROM checksum test

0 = Failed, 1 = Passed