Backlight, Powering off, Navigation – YSI Pro1030 User Manual
Page 7: First power on, System setup menu, Backlight powering off navigation first power on
6
7
BacKlIght
once the instrument is powered on, pressing the power/backlight key
will turn on the display backlight. The backlight will remain on until
the key is pressed again or after two minutes of not pressing any key
on the keypad.
powerIng off
To turn the instrument off, press and hold the power/backlight key
for three seconds.
navIgatIon
The up and down arrow keys allow you to navigate through the
functions of the Pro1030.
NAVIGATING THE RUN SCREEN
When in the Run screen, the up and down arrow keys will move the
highlighted box along the bottom options. once a box is highlighted,
press enter to access the highlighted option.
Description of Run screen box functions from left to right:
Option
Description
SAVE
Highlight and press enter to save displayed
data to memory.
DATA
Highlight and press enter to view and/or erase
saved data.
NAVIGATING THE SYSTEM SETUP MENU
When in the System Setup menu, the up and down arrow keys will
move the highlighted bar up and down the system setup options. See
the System Setup menu section of this manual for more information
about these options.
fIrst power on
The instrument will step through an initial configuration when powered
on for the first time. This will set the language. Use the up or down
arrow keys to highlight the appropriate language, then press enter to
confirm (figure 6).
Figure 6, Select language
After selecting a language, the Run screen will be displayed. The
next time the instrument is powered up, the Run screen will display
immediately after the splash screen.
S Y S T E M S E T U P M E N U
Press the menu
key to access the System Setup menu. The System
Setup menu contains two screens notated as ‘pages’. The current page
is indicated near the bottom of the display (figure 7).
Use the up and down arrow keys to scroll through menu options and
menu pages.
EXITING THE SYSTEM SETUP MENU
To exit the System Setup menu, press the down arrow key until the ESC
- Exit box is highlighted, then press enter to return to the Run screen.