Technical data – YSI 4010-2 pH Adapter DIN / BNC User Manual
Page 5
4010-2/3 pH Adapter DIN / 4010-2/3 pH Adapter BNC
General information
5
ba76144e01
02/2013
Technical data
Note
The latest version of the present operating manual can be found on the Inter-
net under
www.YSI.com
.
General data
Electrodes that can be connected
All pH electrodes with DIN or BNC
plug head connector
Application
characteristics
pH measuring range
Depending on the electrode used
Allowed temperature range
0 ... 80 °C (32 ... 176 °F)
Typical application
Applications with special pH elec-
trodes
Measuring
characteristics of
the IDS measuring
electronics in the
4010-2/3 pH
Adapter
Measured pa-
rameter
Measuring range
Resolution
Accuracy
(± 1 digit)
pH
-2.000 ... 20.000
0.001
± 0.004
U [mV]
-1000.0 ... +1000.0
0.1
± 0.2
T [°C]
0.1
± 0.2
Material, electrical
connection
Material of the plug enclosure
ABS
Electrode connection
DIN or BNC
Meter connection
Mounting slot with contacts
Electrical data of
the electrode
connection
Input resistance
> 5 * 10
12
ohm
Input current
< 1 * 10
-12
A