YSI IQ SensorNet MIQ/JBR Module User Manual
Page 11

MIQ/JBR
Installation
3 - 3
ba76077d01
01/2012
Terminator switch
For the setting of the terminator switches within a signal range, the
same rules apply as for systems without MIQ/JBR (see chapter
I
NSTALLATION
of the IQ S
ENSOR
N
ET
system operating manual).
The longest cable section of each signal range is determined.
On both ends of the longest cable section of a signal range the
terminator switch must be set to On. All other terminator switches must
be set to Off.
Fig. 3-3
Settings of the SN terminator switch
for an IQ S
ENSOR
N
ET
system with 2 MIQ/JBR modules
Note
The proceeding to determine the longest cable section can be found in
chapter I
NSTALLATION
of the IQ S
ENSOR
N
ET
system operating manual.
The proceeding to set the terminator switches can also be found in
chapter I
NSTALLATION
of the IQ S
ENSOR
N
ET
system operating manual.