Troubleshooting, Specifications, Conversions – YSI EcoSense DO200A User Manual
Page 8: Recommended spare parts list
7
TROUBLESHOOTING
Main Display reads:
Possible Solutions:
“ovEr” or “undr”
Check membrane and electrolyte solution.
Clean anode and cathode.
Return product for service.
Secondary Display reads:
Possible Solutions:
“undr”
Heat the sample to above –6.0 °C
Return product for service.
“ovEr”
Cool sample to below 46.0 °C
Return product for service
SPECIFICATIONS
Display Range
Accuracy
Resolution
Dissolved O
2
(ppm or mg/L)
0 to 20.00 ppm (mg/L)
±2 % of the reading or ±2% air
saturation, whichever is greater
0.01 mg/L
Dissolved O
2
% air-sat
0 to 200.0 %
±2% of the reading or ±0.2 ppm,
whichever is greater.
0.1 %
Temperature °C
-6.0 to 46.0 °C
21 to 115 °F
±0.3 °C ±1 digit
0.1 °C
Pressure Compensation
600 to 1100 mBar (450 to 825 mmHg)
Salinity Compensation
From 0.0 to 40.0 ppt
ATC Probe
Thermistor, 10K, at 25C
Calibration Backup
Yes
Audio Feedback
Yes, on all keys
Power Source
One 9V battery
Operating Temperature
0 to 50C (32 to
122 °F)
Instrument Casing
Waterproof IP 67
Weight (with battery)
270 grams (.61 lbs)
Dimensions (L x W x D)
18.7 cm x 7.6 cm x 3.8 cm (7.3 in x 3 in x 1.5 in)
CONVERSIONS
To Convert:
Multiply by:
Inches of Hg to mBar
33.864
Inches of Hg to mmHg
25.4
mmHg to mBar
1.333
RECOMMENDED SPARE PARTS LIST
Item #606035REF
Revision A; July 2012
For the latest version of this manual, visit www.ysi.com
PART #
DESCRIPTION
200-4
4 meter (approx. 13 feet) probe and cable assembly
200-10
10 meter (approx. 33 feet) probe and cable assembly
200-BOD Self-stirring BOD lab probe and cable assembly with power supply
485
DO carrying case, soft sided
606036
DO carrying case, hard sided
5908
Membrane kit, 1.25 mil PE (605306), six cap membranes and KCl solution