Appendix a - specifications – YSI 60 User Manual
Page 30
26
12. Appendix A - Specifications
Materials: ABS, Stainless Steel, and other materials
Dimensions:
Height:
9.5 inches
(24.13 cm)
Thickness:
2.2 inches
(5.6 cm)
Width:
3.5 inches max.
(8.89 cm)
Weight:
1.7 pounds (w/ 10’ cable) (.77 kg)
Display:
2.3”W x 1.5”L
(5.8 cm W x 3.8 cm L)
Power: 6 AA-size Alkaline Batteries (included)
Approximately 100 hours operation from each new set of batteries
Automatic shutoff after 10 hours without a key press
Water Tightness: Meets or exceeds IP65 standards
Probe Operating Environment
Medium: fresh, sea, or polluted water and most other liquid solutions.
Temperature: -5 to +75 °C
Depth: 0 to 10, 0 to 25, 0 to 50, or 0 to 100 feet (depending on cable length)
Meter Ambient Operating/Storage Temperature: -5 to +45 °C
System Performance Specifications
Measurement Range Resolution
Accuracy
pH
0 to 14
0.01 unit
± 0.1 pH unit within ±10°C
of calibration temperature
or
± 0.2 pH unit within ±20°C
of calibration temperature
Temperature
-5 to +75 °C
0.1 °C
± 0.1°C ±1 LSD
pH Response Time: 3 sec for 95% of the change at 25°C
Temperature Response Time: 20 sec for 95% of the change
Temperature Compensation: Automatic
Data Storage: 50 points with ID number.