Zero calibration – Analytical Industries GPR-1800 IS ATEX Trace PPM Oxygen Transmitter User Manual
Page 25

Advanced Instruments, Inc
25
exposure to high O2 concentrations or air is significantly different if a sensor is being installed than if the 
sensor had been in-service at low oxygen levels for more than 1 week. 
Sensor
Calibration at 
Install 
In-service 
Calibration 
PPM 
Fuel 
Cell 
Air to .1% < 1 
min 
Air to 100 PPM < 
5 min 
Air to < 10PPM < 
60 min 
Similar
Less than 45 min
The above times assume the introduction of a zero gas (low level of oxygen in nitrogen) after span 
calibration. 
For optimum accuracy, the O2 concentration of a span gas should be approximate 50-90% of the full scale 
range of analysis or one range above the analysis range, e.g. 20.9% O2 on the 0-25% range. Conversely, if 
the recommended span gas is not available and air calibration is not an option, a span gas of the same full 
scale range and near the anticipated analysis level (approximately 30-50% of full scale) is acceptable with 
the understanding that the accuracy will suffer slightly. 
Use of span gas near 30% of the full scale range of measurements, at the higher end of the range has the 
effect of “expanding the error” as illustrated by Graph A in Example 1 in the Accuracy section above. 
Prematurely initiating the SPAN CALIBRATION function (before the analyzer reading has stabilized) can 
result in erroneous readings as follows: 
When purging an analyzer to lower ranges and calibrating with a span gas: If the oxygen reading reaches 
less than 2% of the intended calibration range, enter the value of the span gas. If the oxygen reading is 
greater than 2% of the calibration range, add the O2 reading to the value of the span gas (the impact of the 
offset on accuracy is minor but the addition allows the oxygen sensor to continue to purge down and avoid 
negative readings after calibration. 
When installing a new oxygen sensor and calibrating with air, allow 2-3 minutes for the sensor to equilibrate 
in ambient air from storage packaging. Failure to do so can introduce error in calibration. 
Zero Calibration
Typical offset from a PPM analyzer is less than 0.5 PPM. Therefore, for most applications, a Zero calibration 
is not required. However, ZERO calibration feature has been provided to allow the user to precisely measure 
oxygen concentration at the very low levels (less than 0.5 PPM). As described below, accomplishing either 
objective places a degree of responsibility on the user. 
Determining the true offset requires the user to wait (see Online Recovery Time section) until the analyzer 
reading is no longer trending downward (best evidenced by a constant horizontal trend on an external 
recording device. 
The zero offset adjustments is limited to 50% of the most sensitive range of the analyzer. At factory, 
analyzer is QC tested to confirm that the maximum offset is less than 50% of the most sensitive range 
available. Should you observe a zero offset more than 50% of the lowest range, check sample system for 
any possible leaks, integrity of the zero gas and assure that the analyzer has been given enough time to 
stabilize on zero gas before initiating the ZERO CALIBRATION. 
Caution: If adequate time is not allowed for the analyzer to establish the true baseline and a ZERO 
calibration is performed, the analyzer will in all probability display a negative reading in the sample mode 
after a certain period of time. If a negative reading is seen, perform ZERO calibration again. 
