8 Troubleshooting

Symptom

Possible Cause

Recommended Action


Reading does not reflect
expected values


Sensor was not calibrated at the pressure,
flow rate and temperature anticipated in
the sample gas stream


Recalibrate the analyzer using the known oxygen
content of ambient air, 20.9%, and if the sensor
responds we should presume it is correct.


Oxygen reading drifts
toward zero or significant
number of turns of the
span control adjustment is
required to calibrate the
analyzer.


Indication sensor is nearing the end of its
useful life


Replace sensor


Slow response time

Erratic oxygen reading

No oxygen reading


Liquid covering sensing membrane

Presence of interference gases.

Unauthorized maintenance

Defective electrical connection

Sensor failure


Gently remove with alcohol and lint free towel.

Consult factory, replace sensor





Use voltmeter and determine uA or mV output
and contact factory.


High oxygen reading


Inadequate control of pressure and
flowrate

Abnormality in span gas


See Installation Considerations, section 5



Qualify source by recalibrating the analyzer with
the known oxygen content of ambient air, 20.9%