beautypg.com

3B Scientific Critical Point Apparatus User Manual

Page 4

background image

4

the corresponding pressures p

0

and p

1

of the trapped

air, we get:

1

1

0

0

s

p

s

p

=

(2)

Substituting

s

s

s

Δ

+

=

1

0

and rearranging gives:

s

p

p

p

s

Δ

=

0

1

0

1

(3)

Rough calibration of scales:

Open the regulating valve wide.

Loosen the grub screw for the vernier scale by
half a turn (it is now possible to turn the scale
easily on the threaded axle without moving the
handwheel, although a counterpressure acts
against this independent movement).

Wind the handwheel out till you detect a notice-
able resistance.

Without turning the handwheel, turn the vernier
scale on the threaded axle till the 0.0 mark is on
the top and the fixed scale shows approx. 48 mm.

Loosen the knurled screws of the fixed scale and
shift the scale to the side till the 48-mm bar is ex-
actly above the centre line of the vernier scale
(see Fig. 2).

Tighten the knurled screws again. In doing so,
make sure that the fixed scale does not press
against the vernier scale.

0 0

1 9

1 8

1 7

1 6

1 5

10

0

20

30

40

50mm

Fig. 2: Piston position reading at 48.0 mm

Zero correction:

Shut the regulating valve (the pressure in the
measuring cell now corresponds to the ambient
pressure p

0

= 1 bar. To within the accuracy of the

measurement, the manometer should display an
excess pressure of 0 bar).

Wind the handwheel in till an excess pressure of
15

bar has been reached (absolute pressure

p

1

= 16 bar).

Read the piston position s

1

and calculate the

displacement

Δs = s

0

s

1

.

Calculate the zero corrected piston position s

1, corr

using Equation 3.

Adjust the vernier scale to the corrected value
and, if necessary, move the scale again.

If required, wind the handwheel out a little and
secure the vernier scale with the grub screw.

Measurement example:

p

0

= 1 bar, p

1

= 16 bar, p

1

p

0

= 15 bar

s

0

= 48.0 mm, s

1

= 3.5 mm,

Δs = 44.5 mm

Therefore, s

1, corr

= 2.97 mm.

The vernier scale must therefore be adjusted so that
now only 2.97 mm are shown instead of 3.50 mm.

Note:

After calibrating the zero point, it is possible to obtain
qualitatively accurate measured values. With regard
to temperature T and pressure p, it is also possible to
obtain quantitatively accurate measurements of the
isotherms in range around to the critical point where
the two phases exist simultaneously. However, espe-
cially in the liquid phase, the measured isotherms are
rather too widely separated.

6.3 Detailed calibration:

The exact relation between the volume V

G

in the

measuring cell and the scale reading s is dependent
on the volume of oil in the oil chamber. The oil
chamber also expands marginally in proportion to the
pressure as a result of the spring in the manometer
tube. Additionally, when the temperature is in-
creased, the castor oil expands to a greater extent
than the rest of the equipment. This means that the
pressure rises at a slightly greater rate at higher tem-
peratures. All of these phenomena can be calculated
if appropriate calibration has been effected using air
as an ideal gas.

The ideal gas equation would thus be:

R

n

T

V

p

=

(4)

with

mol

K

J

314

8.

R

=

After taking the overpressure reading p

e

, the absolute

pressure can be calculated from:

p

= p

e

+ 1 bar

(6)

The absolute temperature is given by:

T

=

ϑ + ϑ

0

where

ϑ

0

= 273.15°C

(7)

The volume is given by:

s

A

V

=

G

(8)

where

2

cm

14

3,

A

=

and s is the “effective” piston

displacement.

From the measured displacement s

e

, it is possible to

calculate the effective piston displacement as follows: