beautypg.com

3 parameter adjustment - quick setup – VEGA VEGABAR 87 Profibus PA - Operating Instructions User Manual

Page 29

background image

29

6 Set up with the display and adjustment module

VEGABAR 87 • Profibus PA

45047-EN-131107

– Confirm selected menu

– Edit parameter

– Save value

[-] key:

– Presentation, change measured value

– Select list entry

– Select editing position

[+] key:

– Change value of the parameter

[ESC] key:

– Interrupt input

– Jump to next higher menu

The device is adjusted via the four keys of the display and adjustment

module. The LC display indicates the individual menu items. The

functions of the individual keys are shown in the above illustration. Ap-

prox. 60 minutes after the last pressing of a key, an automatic reset to

measured value indication is triggered. Any values not confirmed with

[OK] will not be saved.

6.3 Parameter adjustment - Quick setup

To quickly and easily adapt the sensor to the application, select

the menu item "Quick setup" in the start graphic on the display and

adjustment module.

Carry out the following steps in the below sequence.
You can find "Extended adjustment" in the next sub-chapter.
1. Sensor address

In the first menu item you have to enter a sensor address. The selec-

tion switches on the electronics module are preset to sensor address

126. This means that the sensor address can be changed via the

display and adjustment module.

2. Measurement loop name

In this menu item you assign a suitable measurement loop name.

Permitted are names with max. 19 characters.
3. Application

In this menu item, you activate/deactivate the slave for the electronic

differential pressure and select the application. The application com-

prises process pressure and level measurement.

Adjustment system

Quick setup - Presettings