7 supplementary electronics, 8 switch-on phase – VEGA VEGAFLEX 82 4 … 20 mA_HART two-wire User Manual
Page 28
28
5 Connecting to power supply
VEGAFLEX 82 • 4 … 20 mA/HART two-wire
41829-EN-130612
5.7 Supplementary electronics
You can use the supplementary electronics - second current output to
provided a second measured value.
Both current outputs are passive and need a power supply.
1
2
+
( )
(-)
I
4...20mA
I I
4...20mA
7
8
+
( )
(-)
3
1
2
Fig. 23: Connection compartment double chamber housing, supplementary
electronics - second current output
1 First current output (I) - Voltage supply and signal output (HART)
2 Second current output (II) - Voltage supply and signal output (without HART)
3 Ground terminal for connection of the cable screen
The radio module PLICSMOBILE is an external GSM/GPRS radio
unit for transmission of measured values and for remote parameter
adjustment.
1
Bus
USB
Status
Test
SIM-Card
1
2
+
( )
(-)
Fig. 24: Radio module PLICSMOBILE integrated in the connection compart-
ment
1 Voltage supply
You can find detailed information on connection in the supplementary
instructions "PLICSMOBILE GSM/GPRS radio module".
5.8 Switch-on phase
After connecting the instrument to power supply or after a voltage
recurrence, the instrument carries out a self-check for approx. 30 s:
•
Internal check of the electronics
•
Indication of the instrument type, hardware and software version,
measurement loop name on the display or PC
•
Indication of the status message "F 105 Determine measured
value" on the display or PC
•
The output signal jumps to the set error current
Supplementary electron-
ics - Second current
output
Supplementary elec-
tronics - Radio module
PLICSMOBILE