VEGA VEGAFLEX 86 Profibus PA User Manual
Page 92
92
11 Supplement
VEGAFLEX 86 • Profibus PA
44232-EN-130910
Protection rating
Ʋ unassembled
IP 20
Ʋ mounted into the housing without
cover
IP 40
Materials
Ʋ Housing
ABS
Ʋ Inspection window
Polyester foil
Integrated clock
Date format
Day.Month.Year
Time format
12 h/24 h
Time zone Ex factory
CET
Rate deviation max.
10.5 min/year
Measurement electronics temerature
Resolution
1 °C (1.8 °F)
Accuracy
±1 °C (1.8 °F)
Voltage supply
Operating voltage
Ʋ Non-Ex instrument
9 … 32 V DC
Ʋ EEx-ia instrument - Power supply
FISCO model
9 … 17.5 V DC
Ʋ EEx-ia instrument - Power supply
ENTITY model
9 … 24 V DC
Ʋ EEx-d instrument
14 … 32 V DC
Number of sensors per DP/PA segment coupler, max.
Ʋ Non-Ex
32
Ʋ Ex
10
Electrical protective measures
Protection, depending on housing version
Ʋ Plastic housing
IP 66/IP 67
Ʋ Aluminium housing, stainless steel
housing - investment casting, stain-
less steel housing - electro-polished
IP 66/IP 68 (0.2 bar)
5)
Ʋ Aluminium and stainless housing, in-
vestment casting (optionally available)
IP 66/IP 68 (1 bar)
Overvoltage category
III
Protection class
III
Approvals
Instruments with approvals can have different technical data depending on the version.
5)
The prerequisites for maintaining the protection rating are a suitable cable as well as correct mounting.