6 Set up the sensor with the indicating and adjustment module
VEGAPULS 62 standpipe version • Modbus and Levelmaster protocol
41363-EN-121011
With radar sensors with standpipe antenna, the sensor length is
already preset in this menu item. When shortening the standpipe
antenna afterwards, this value must be appropriately corrected.
When a reset is carried out, all settings (with only a few exceptions)
are reset. The exceptions are: PIN, language and lighting.
The following reset functions are available:
Delivery status: Restoring the parameter settings at the time of ship-
ment from the factory incl. the order-specific settings. A created false
signal suppression, user-programmable linearization curve as well as
the measured value memory will be deleted.
Basic settings: Resetting the parameter settings incl. special and
laboratory parameters to the default values of the respective instru-
ment. A created false signal suppression, user programmable lineari-
zation curve as well as the measured value memory will be deleted.
Setup: Resetting of the parameter settings to the default values of the
respective instrument in the menu item Setup. Order-related settings
remain but are not taken over into the current parameters. User-
generated false signal suppression, user-programmed linearization
curve, measured value memory as well as event memory remain
untouched. The linearization is set to linear.
False signal suppression: Deleting a previously created false signal
suppression. The false signal suppression created in the factory
remains active.
Peak values distance: Resetting the measured min. and max. dis-
tances to the actual measured value.
Select the requested reset function [->] and confirm with [OK].
The following table shows the default values of VEGAPULS 62:
Menu section
Menu item
Default value
Setup
Measurement
loop name
Sensor
Min. adjustment
Length of the standpipe Ex factory
Damping
0.0 s
Display
Language
Like order
Displayed value
Distance
Display unit
m(d)
Scaling
0.00 %, 0 l
100.00 %, 100 l
Additional adjustments/
Probe length
Additional adjustments/
Reset