beautypg.com

5 setup, 1 setup – VEGA VEGABAR series 80 User Manual

Page 10

background image

10

5 Setup

Electronics module • VEGABAR series 80

45054-EN-131119

5 Setup

5.1 Setup

After mounting a new electronics module and connecting to power

supply, the following data are loaded automatically from the meas. cell

electronics:

Instrument type

Serial number

Date of manufacture

Measuring range

Default

The instrument is generally ready for operation.

If you are using an electronics module without sensor serial number

or a "universal electronics module" from stock, you can load the

order-specific data into the sensor on site if required.
Information:

When loading on site, the file must first be downloaded from the Inter-

net (see "Load sensor data").
Go at www.vega.com to "Instrument search". The order data of the

sensor are displayed after entering the serial number.
Below the order data you will find "Sensor data for Service-DTM" as

XML file. Load this file with "Save as" on your PC and transfer it then

via PACTware and the Service DTM to the sensor.

As a rule, settings already carried out on site with the previous elec-

tronics module, such as min./max. adjustment etc., must be repeated.

Tip:

To do this, use the copy function of the display and adjustment mod-

ule or the adjustment software PACTware.

Load order-specific sen-

sor data

Adjustment