VEGA VEGAFLEX 81 4 … 20 mA_HART - four-wire Coax probe User Manual
Page 57
57
9 Diagnostics and service
VEGAFLEX 81 • 4 … 20 mA/HART - four-wire
42280-EN-130314
Code
Text mes-
sage
Cause
Rectification
F260
Error in the
calibration
– Error in the calibration car-
ried out in the factory
– Error in the EEPROM
– Exchanging the electronics
– Send instrument for repair
F261
Error in the
instrument
settings
– Error during setup
– Error when carrying out a
reset
– False signal suppression
faulty
– Carry out a reset
– Repeat setup
F264
Installation/
Setup error
– Error during setup
– Check or correct installation
and/or parameter adjust-
ment
– Check probe length
F265
Measurement
function dis-
turbed
– Sensor no longer carries
out a measurement
– Carry out a reset
– Separate operating voltage
briefly
F266
Impermissi-
ble operating
voltage
– Operating voltage below
specified range
– Check electrical connection
– if necessary, increase
operating voltage
F267
No executable
sensor soft-
ware
– Sensor cannot start
– Exchanging the electronics
– Send instrument for repair
The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes
as well as corrective measures.
Code
Text mes-
sage
Cause
Rectification
C700
Simulation ac-
tive
– A simulation is active
– Finish simulation
– Wait for the automatic end
after 60 mins.
The following table shows the error codes and text messages in the
status message "Out of specification" and provides information on
causes as well as corrective measures.
Code
Text mes-
sage
Cause
Rectification
S600
Unpermissi-
ble electronics
temperature
– Temperature of the pro-
cessing electronics in the
non-specified section
– Check ambient temperature
– Isolate electronics
– Use instrument with higher
temperature range
Function check
Out of specification