4 saving the parameter adjustment data – VEGA VEGAFLEX 81 4 … 20 mA_HART - four-wire Coax probe User Manual
Page 52
52
7 Setup with PACTware
VEGAFLEX 81 • 4 … 20 mA/HART - four-wire
42280-EN-130314
instrument parameter adjustment. The complete data set is down-
loaded from the device. This procedure can last several minutes.
Create instrument documentation?
This function is used to print or create a PDF file of the current param-
eter adjustment. To read the PDF file, you need a suitable program
(for example Acrobat Reader). To print or create the PDF file, all data
are downloaded from the device. For this function, the full version
of the DTM Collection is required. This procedure can last several
minutes.
Store echo curve of the setup in the sensor?
Have you completed the initial setup of the instrument? In such case,
we recommend storing the current signal conditions in the device for
later instrument tests and diagnostics.
Lock adjustment with PIN after setting?
The instrument is locked with the current PIN. A parameter adjust-
ment is possible only after the PIN is entered again.
7.4 Saving the parameter adjustment data
We recommend documenting or saving the parameter adjustment
data via PACTware. That way the data are available for multiple use or
service purposes.