VEGA VEGAFLEX 83 4 … 20 mA_HART - four-wire Polished version User Manual
Page 52
52
7 Setup with PACTware
VEGAFLEX 83 • 4 … 20 mA/HART - four-wire
41840-EN-130314
Extended adjustment
With the extended adjustment, you carry out the parameter adjust-
ment for the instrument via the clear menu structure in the DTM
(Device Type Manager). This enables additional and special settings
over and above those offered by quick setup.
Maintenance
Under the menu item "Maintenance" you get comprehensive and
important support for servicing and maintenance. You can call up
diagnostic functions and carry out an electronics exchange or a
software update.
Click to the button "Quick setup", to start the assistant-driven adjust-
ment for a simplified and reliable setup.
Device name
Here, you can find the instrument name. You cannot change this line
because the instrument name is unmodifiably saved in the instrument.
Serial number
Here, you can find the serial number of your instrument. You cannot
change this line because the serial number is unmodifiably saved in
the instrument.
Measurement loop name
Here you can enter a suitable measurement loop name for your VE-
GAFLEX 83. You can enter a name with max. 19 characters. You can
use capital and small letters as well as numbers. The following special
characters are also possible: + - . : , ( ) / < >
Probe length modified?
If you have modified the probe length, this must be entered in the
selection field.
•
If you select "No", then the instrument uses automatically the
preset length of the default setting.
•
If you select "Yes", then you can enter in another field the modified
length of the instrument.
Probe length L from seal surface
If you have modified the length of the probe, you can enter in this field
the modified probe length. Keep the selected unit in mind.
Determine probe length automatically
If you do not know the probe length, you can have the length of the
probe determined automatically. The requirement for this is a probe
unrestricted and not covered by the medium.
Click to "Carry out now", to start the automatic length determination.
Product type
Here you can see which medium type your instrument is suitable for.
If this function is inactive, the medium type your instrument is suitable
for was already preset.
Start quick setup
Step 1 of 7
Instrument configuration
Step 2 of 7
Application