64

11 Supplement

VEGAPULS 66 enamel • 4 … 20 mA/HART - two-wire

36527-EN-121011

Vessel pressure relating to the antenna

system

1 … 40 bar/-100 … 4000 kPa/-(-14.5 … 580.2 psig)

Vessel pressure relating to the flange

nominal stage

see supplementary instructions manual "Flanges ac-

cording to DIN-EN-ASME-JIS"

Vibration resistance

4 g at 5 … 200 Hz according to EN60068-2-6 (vibration

with resonance)

Data on rinsing air connection
Pressure max.

6 bar (87.02 psi)

Air flow without reflux valve

Ʋ 0.1 bar (1.45 psi)

4.2 m

3

/h

Ʋ 1.2 bar (7.25 psi)

10.2 m

3

/h

Air flow with reflux valve

Ʋ 0.2 bar (2.9 psi)

2.0 m

3

/h

Ʋ 0.55 bar (7.98 psi)

4.2 m

3

/h

Ʋ 1.2 bar (17.4 psi)

7.2 m

3

/h

Thread

G¼ A

Closure

Ʋ with non-Ex

Dust protection cover of PE

Ʋ with Ex

Threaded plug of 316Ti

2 reflux valves - enclosed unassembled (with non-Ex optional, with Ex included in the scope of

delivery)

Ʋ Material

316Ti

Ʋ Seal

FKM (Viton), FFKM (Kalrez 6375)

Ʋ for tube diameter

10 mm

Ʋ opening pressure

0.5 bar (7.252 psi)

Ʋ Nominal pressure stage

PN 250

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry

Ʋ Cable gland

M20 x 1.5 (cable: ø 5 … 9 mm)

Ʋ Cable entry

½ NPT

Ʋ Blind stopper

M20 x 1.5; ½ NPT

Ʋ Closing cap

M20 x 1.5; ½ NPT

Plug options

Ʋ Signal circuit

M12 x 1, according to ISO 4400, Harting HAN, 7/8" FF

Ʋ Indication circuit

M12 x 1

Wire cross-section (spring-loaded terminals)

Ʋ Massive wire, cord

0.2 … 2.5 mm² (AWG 24 … 14)

Ʋ Stranded wire with end sleeve

0.2 … 1.5 mm² (AWG 24 … 16)

Electromechanical data - version IP 66/IP 68 (1 bar)
Options of the cable entry