beautypg.com

VEGA VEGABAR 52 4 … 20 mA_HART - climate compensated User Manual

Page 52

background image

52

10 Supplement

VEGABAR 52 • 4 … 20 mA/HART - climate compensated

36718-EN-130321

Nominal measuring ranges and overload capability in psig
The specifications are only an overview and refer to the measuring cell. Limitations due to the ma-

terial and version of the process fitting are possible. The specifications on the nameplate apply.

Nominal range

Overload capacity, max.

pressure

Overload capacity, min. pres-

sure

Gauge pressure
0 … +3 psig

+500 psig

-15 psig

0 … +6 psig

+700 psig

-15 psig

0 … +15 psig

+700 psig

-15 psig

0 … +35 psig

+725 psig

-15 psig

0 … +70 psig

+950 psig

-15 psig

0 … +150 psig

+1300 psig

-15 psig

0 … +350 psig

+1900 psig

-15 psig

0 … +900 psig

+2900 psig

-15 psig

-15 … 0 psig

+500 psig

-15 psig

-15 … +25 psig

+700 psig

-15 psig

-15 … +70 psig

+950 psig

-15 psig

-15 … +150 psig

+1300 psig

-15 psig

-15 … +350 psig

+1900 psig

-15 psig

-15 … +900 psig

+2900 psig

-15 psig

-1.5 … +1.5 psig

+500 psig

-15 psig

-3 … +3 psig

+500 psig

-12 psig

-7 … +7 psig

+700 psig

-15 psig

Reference conditions and actuating variables (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1

Ʋ Temperature

+15 … +25 °C (+59 … +77 °F)

Ʋ Relative humidity

45 … 75 %

Ʋ Air pressure

860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)

Determination of characteristics

Limit point adjustment according to IEC 61298-2

Characterstic curve

Linear

Reference installation position

upright, diaphragm points downward

Influence of the installation position

< 0.2 mbar/20 Pa (0.003 psig)

Deviation determined according to the limit point method according to IEC 60770

11)

Applies to the digital signal output (HART, Profibus PA, Foundation Fieldbus) as well as to ana-

logue current output 4 … 20 mA and refers to the set span. Turn down (TD) is the ratio nominal

measuring range/set span.
Deviation with absolutely flush process fittings EV, FT

Ʋ Turn down 1 : 1 up to 5 : 1

< 0.05 %

11)

Incl. non-linearity, hysteresis and non-repeatability.