beautypg.com

10 supplement, 1 technical data – VEGA VEGABAR 52 4 … 20 mA_HART - climate compensated User Manual

Page 49

background image

49

10 Supplement

VEGABAR 52 • 4 … 20 mA/HART - climate compensated

36718-EN-130321

10 Supplement

10.1 Technical data

General data
Parameter, pressure

Gauge pressure, vacuum

Measuring principle

Ceramic-capacitive, dry measuring cell

Communication interface

I²C bus

Materials and weights
Material 316L corresponds to 1.4404 or 1.4435
Materials, wetted parts

Ʋ Process fitting

316L, PVDF, Alloy C-22, Alloy C-276, Duplex 1.4462,

Titanium Grade 2

Ʋ Diaphragm

sapphire ceramic

®

(99.9 % oxide ceramic)

Ʋ Joining material diaphragm/Basic ele-

ment measuring cell

Glass solder

Ʋ Measuring cell seal

FKM (VP2/A, A+P70.16), EPDM (A+P 75.5/KW75F),

FFKM (Kalrez 6375, Perlast G75S, Perlast G75B)

Material seal process fitting

Ʋ Thread G½ (EN 837)

Klingersil C-4400

Ʋ Thread G1½ (DIN 3852-A)

Klingersil C-4400

Ʋ M44 x 1.25 (DIN 13)

FKM, FFKM, EPDM

Surface quality hygienic fittings, typ.

R

a

< 0.8 µm

Ʋ Surface quality, typ.

Materials, non-wetted parts

Ʋ Electronics housing

Plastic PBT (polyester), Alu die-casting powder-coated,

316L

Ʋ External housing

plastic PBT (Polyester), 316L

Ʋ Socket, wall mounting plate external

housing

plastic PBT (Polyester), 316L

Ʋ Seal between socket and wall mount-

ing plate

EPDM (fixed connected)

Ʋ Seal below wall mounting plate

EPDM (only with 3A approval)

Ʋ Seal, housing cover

NBR (stainless steel housing), silicone (Alu/plastic hous-

ing)

Ʋ Inspection window in housing cover

for display and adjustment module

Polycarbonate (UL-746-C listed)

Ʋ Ground terminal

316Ti/316L

Ʋ Ohmic contact

Between ground terminal and process fitting

Ʋ Connection cable between transmitter

and external electronics housing with

IP 68 version

PUR

Ʋ Type label support on connection

cable

PE hard