beautypg.com

3 dimensions – VEGA VEGABAR 53 Foundation Fieldbus User Manual

Page 56

background image

56

10 Supplement

VEGABAR 53 • Foundation Fieldbus

36724-EN-130321

– Highest calibrated value. For calibration of the high limit point you give the high measurement

value (pressure) to the sensor and transfer this point as HIGH to the transmitter. Unit derives

from 'Sensor_range.unit'

CAL_POINT_LO

– Lowest calibrated value. For calibration of the low limit point you give the low measurement

value (pressure) to the sensor and transfer this point as LOW to the transmitter. Unit derives

from 'Sensor_range.unit'

CAL_MIN_SPAN

– Minimum calibration span value allowed. Necessary to ensure that when calibration is done,

the two calibrated points (high and low) are not too close together. Unit derives from 'Sen-

sor_range.unit'

SCALE_IN

– Min/max-adjustment: Upper and lower calibrated points of the sensor. Unit derives from 'Sen-

sor_range.unit'

trimmed_value

– Sensor value after the trim processing. Unit derives from 'Sensor_range.unit'

sensor_sn

– Sensor serial number

temperature

– Process temperature. Selected as input to AIFB by setting 'Channel' = 4. Unit derives from

'Temperature.unit'

temperature_unit

– Unit code of 'Temperature', 'Max/Min_peak_temperature_value'

– °C, °F, K, °R

max_peak_temperature_value

– Holds the maximum process temperature. Write access resets to current value. Unit derives

from 'Temperature.unit'

– Write access resets to current value

min_peak_temperature_value

– Holds the minimum process temperature. Write access resets to current value. Unit derives

from 'Temperature.unit'

– Write access resets to current value

10.3 Dimensions

The two chamber housings are not available with instruments with 4 … 20 mA signal output

Housing in IP 66/IP 67

2

1

ø 86 mm (3.39")

~ 116 mm (4.57")

116 mm (4.57"

)

M20x1,5

M20x1,5/

½ NPT

~ 87 mm (3.43")

M16x1,5

ø 86 mm

(3.39")

120 mm (4.72"

)

M20x1,5/

½ NPT

Fig. 30: Housing in protection IP 66/IP 67, with integrated display and adjustment module the housing is 9 mm

(

1

/

64

") higher