9 Diagnosis, Asset Management and service
VEGAPULS 66 standpipe version • Foundation Fieldbus
36526-EN-121011
Code
Text mes-
sage
Cause
Rectification
PA DevS-
pec
Diagnosis
F260
Error in the
calibration
– Error in the calibra-
tion carried out in the
factory
– Error in the EEPROM
– Exchanging the elec-
tronics
– Send instrument for
repair
Bit 9
F261
Error in the
configura-
tion
– Error during setup
– False signal suppres-
sion faulty
– Error when carrying
out a reset
– Repeat setup
– Repeat reset
Bit 10
F264
Installa-
tion/Setup
error
– Adjustment not within
the vessel height/
measuring range
– Max. measuring range
of the instrument not
sufficient
– Check or correct
installation and/or
parameter adjustment
– Use an instrument
with bigger measuring
range
Bit 11
F265
Meas-
urement
function
disturbed
– Sensor no longer
carries out a measure-
ment
– Operating voltage
too low
– Check operating
voltage
– Carry out a reset
– Separate operating
voltage briefly
Bit 12
The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes
as well as corrective measures.
Code
Text mes-
sage
Cause
Rectification
C700
Simulation ac-
tive
– A simulation is active
– Finish simulation
– Wait for the automatic end
after 60 mins.
The following table shows the error codes and text messages in the
status message "Out of specification" and provides information on
causes as well as corrective measures.
Code
Text mes-
sage
Cause
Rectification
S600
Unpermissi-
ble electronics
temperature
– Temperature of the elec-
tronics in the non-specified
section
– Check ambient temperature
– Isolate electronics
– Use instrument with higher
temperature range
S601
Overfilling
– Danger of vessel overfilling
– Make sure that there is no
further filling
– Check level in the vessel
Function check
Out of specification