beautypg.com

VEGA VEGABAR 66 Foundation Fieldbus User Manual

Page 56

background image

56

10 Supplement

VEGABAR 66 • Foundation Fieldbus

36740-EN-130321

Product temperature, suspension cable/measuring cell seal

Ʋ PE/FKM

-20 … +60 °C (-4 … +140 °F)

Ʋ PE/EPDM

-20 … +60 °C (-4 … +140 °F)

Ʋ PUR/FKM

-20 … +80 °C (-4 … +176 °F)

Ʋ PUR/EPDM

-20 … +80 °C (-4 … +176 °F)

Ʋ FEP/FKM

-20 … +100 °C (-4 … +212 °F)

Ʋ FEP/FFKM

-10 … +100 °C (+14 … +212 °F)

Ʋ FEP/EPDM

-20 … +100 °C (-4 … +212 °F)

Product temperature, connection tube/seal meas. cell

Ʋ FKM

-20 … +100 °C (-4 … +212 °F)

Ʋ FFKM

-10 … +100 °C (+14 … +212 °F)

Ʋ EPDM

-20 … +100 °C (-4 … +212 °F)

Product temperature, transmitter protection/seal meas. cell

Ʋ PVDF/FKM or EPDM

-20 … +60 °C (-4 … +140 °F)

Ʋ PVDF/FFKM

-10 … +60 °C (+14 … +140 °F)

Ʋ PE/FKM or EPDM

-20 … +60 °C (-4 … +140 °F)

Calibration position

upright, diaphragm points downward

Influence of the installation position

< 0.2 mbar/20 Pa (0.003 psig)

Vibration resistance

mechanical vibrations with 4 g and 5 … 100 Hz

15)

With version with connection tube, vibration resistance 1 g. With lengths > 0.5 m (1.64 ft) the tube

must be supported in addition.
Shock resistance

Acceleration 100 g/6 ms

16)

Electromechanical data - version IP 66/IP 67
Cable entry/plug

17)

Ʋ Single chamber housing

– 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x

blind plug M20 x 1.5

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT

or:

– 1␣x plug (depending on the version), 1␣x blind stopper

M20␣x␣1.5

or:

– 2 x blind plug M20 x 1,5

15)

Tested according to the guidelines of German Lloyd, GL directive 2.

16)

Tested according to EN 60068-2-27.

17)

Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.