6 Set up with the indicating and adjustment module

VEGAPULS 62 standpipe version • 4 … 20 mA/HART - four-wire

36508-EN-121011

Menu

Menu item

Default value

Lock operation

Released

Display

Language

Like order

Displayed value

Distance

Display unit

m

Scaling size

Volume
l

Scaling

0.00 lin %, 0 l
100.00 lin %, 100 l

Backlight

Switched off

Additional adjust-

ments

Distance unit

m

Temperature unit °C
Probe length

Length of the standpipe Ex factory

Linearisation

curve

Linear

HART mode

Standard
Address 0

6.4 Saving the parameter adjustment data

We recommended noting the adjusted data, e.g. in this operating

instructions manual, and archiving them afterwards. They are thus

available for multiple use or service purposes.
If the instrument is equipped with an indicating and adjustment

module, the data in the sensor can be saved in the indicating and

adjustment module. The procedure is described in the operating

instructions manual "Indicating and adjustment module" in the menu

item "Copy sensor data". The data remain there permanently even if

the sensor power supply fails.
The following data or settings for adjustment of the indicating and

adjustment module are saved:

All data of the menu "Setup" and "Display"

In the menu "Additional adjustments" the items "Sensor-specific

units, temperature unit and linearization"

The values of the user programmable linearisation curve

The function can also be used to transfer settings from one instru-

ment to another instrument of the same type. If it is necessary to

exchange a sensor, the indicating and adjustment module is inserted

into the replacement instrument and the data are likewise written into

the sensor via the menu item "Copy sensor data".