Measurement error during filling – VEGA VEGAPULS 63 (≥ 2.0.0 - ≥ 4.0.0) 4 … 20 mA_HART two-wire User Manual
Page 55
55
9 Diagnosis, Asset Management and service
VEGAPULS 63 • 4 … 20 mA/HART two-wire
36511-EN-130228
Measurement error with constant level
Fault description
Error pattern
Cause
Rectification
1. Measured value
shows a too low or too
high level
Level
time
0
– Min./max. adjustment not
correct
– Adapt min./max. adjustment
– Wrong linearization curve
– Adapt linearization curve
– Installation in a bypass tube or
standpipe, hence running time
error (small measurement error
close to 100 %/large error close
to 0 %)
– Check parameter "Application"
with respect to vessel form,
adapt if necessary (bypass,
standpipe, diameter)
2. Measured value
jumps towards 0 %
Level
time
0
– Multiple echo (vessel top,
product surface) with amplitude
higher than the level echo
– Check parameter "Application",
especially vessel top, product
type, dished end, high dielectric
figure, adapt if necessary
3. Measured value
jumps towards 100 %
Level
time
0
– Due to the process, the ampli-
tude of the product echo sinks
– A false signal suppression was
not carried out
– Carry out false signal suppres-
sion
– Amplitude or position of a
false echo has changed (e.g.
condensation, buildup); false
signal suppression no longer
matches
– Determine the reason for the
changed false signals, carry out
false signal suppression, e.g.
with condensation
Measurement error during filling
Fault description
Error pattern
Cause
Rectification
4. Measured value re-
mains unchanged
during filling
Level
time
0
– False echoes in the close range
too big or product echo too
small
– Strong foam or spout genera-
tion
– Max. adjustment not correct
– Eliminate false echoes in the
close range
– Check measurement situation:
Antenna must protrude out of
the socket, installations
– Remove contamination on the
antenna
– Minimize interfering instal-
lations in the close range by
changing the polarization
direction
– Create a new false signal sup-
pression
– Adapt max. adjustment
5. Measured value re-
mains in the bottom
section during filling
Level
time
0
– Echo from the tank bottom
larger than the product echo,
for example, with products with
ε
r
< 2.5 oil-based, solvents
– Check application parameters
Medium, Vessel height and
Floor form, adapt if necessary
6. Measured value re-
mains momentarily
unchanged during fill-
ing and then jumps to
the correct level
Level
time
0
– Turbulence on the product
surface, quick filling
– Check application parameters,
change if necessary, e.g. in
dosing vessel, reactor