6 Set up with the display and adjustment module
SOLITRAC 31 • 4 … 20 mA/HART - four-wire
40090-EN-131119
Information:
In this operating instructions manual, the instrument-specific param-
eters are described. Further general parameters are described in the
operating instructions manual "Display and adjustment module".
Caution:
During the first setup or after an instrument reset the instrument starts
with preset standard values. These value are not suitable for your ap-
plication and must be replaced by real values.
Carry out a setup in the sequence described in the following.
The main menu is divided into five sections with the following func-
tions:
Setup: Settings, e.g. for measurement loop name, isotope, applica-
tion, background radiation, adjustment, signal output
Display: Settings, for example language, measured value display
Diagnosis: Information, for example, of device status, peak value,
simulation
Additional adjustments: Instrument unit, reset, date/time, copying
function
Info: Instrument name, hardware and software version, date of manu-
facture, instrument features
Check if the correct language is already set for the display. If not, you
can change the language in the menu item "Display/Language".
Start with the setup of SOLITRAC 31.
In the main menu point "Setup", the individual submenu points
should be selected one after the other and provided with the correct
parameters to ensure optimum adjustment of the measurement. The
procedure is described in the following.
Stick with the normal sequence of the menu items as closely as pos-
sible.
Setup
In this menu item you can assign an unambiguous name to the sen-
sor or the measurement loop.
This parameter is described in the operating instructions manual
"Display and adjustment module".
In this menu item you can adjust the SOLITRAC 31 to the integrated
isotope in the source container.
Instrument start
Main menu
Procedure
Setup - Measurement
loop name
Setup/Isotope