beautypg.com

VEGA VEGAFLEX 66 (-200…+400°C) 4 … 20 mA_HART four-wire User Manual

Page 51

background image

L

0

-3 mm (-0.118

")

3 mm (0.118

")

-10 mm (-0.394

")

5 mm (0.197

")

-5 mm (-0.197

")

0,15 m (5.906

"

)

0,06 m (2.362

"

)

0,05 m

(1.968

")

0,1 m

(3.937

")

1

1

Fig. 25: Deviation VEGAFLEX 66 in coax version in water

1

Dead zone - no measurement possible in this area

L

Probe length

L

0

-3 mm (-0.118

")

3 mm (0.118

")

-10 mm (-0.394

")

10 mm (0.394

")

0,1 m (3.937

"

)

0,15 m

(1.968

")

0,1 m

(3.937

")

0,25 m (9.842

"

)

1

1

Fig. 26: Deviation VEGAFLEX 66 in coax version in oil

1

Dead zone - no measurement possible in this area

L

Probe length

Influence of the ambient temperature to the sensor electronics

T

emperature drift

0.03 %/10 K

relating to the max. measuring range

or max. 0.3 %

T

emperature drift - Digital output

3

mm/10 K relating to the max. measuring range or

max. 10 mm

Ambient conditions

A

mbient, storage and transport temperature

-

40 … +70 °C (-40 … +158 °F)

VEGAFLEX

66 • 4 … 20

mA/HART four-wire

51

10 S

upplement

34174

-

EN

-100426