beautypg.com

VEGA VEGAFLEX 63 Profibus PA User Manual

Page 36

background image

Error

Cause

Removal

In Simatic S7 the
measured value is
always presented
as 0

Only four bytes
are consistently
loaded in the PLC

Use function component SFC 14 to
load 5 bytes consistently

Measured value
on the indicating
and adjustment
module does not
correspond to the
value in the PLC

The menu item
"

Display - Dis-

play value

"

is

not set to "

PA-

Out

"

Check values and correct, if neces-
sary

No connection
between PLC and
PA network

Incorrect adjust-
ment of the bus
parameter and
the segment
coupler-depen-
dent baud rate

Check data and correct, if necessary

Instrument does
not appear during
connection setup

Profibus DP cab-
le pole-reversed

Check cable and correct, if necessary

Incorrect termina-
tion

Check termination at the beginning
and end points of the bus and termin-
ate, if necessary, according to the
specification

Instrument not
connected to the
segment, double
assignment of an
address

Check and correct, if necessary

In Ex applications, the regulations for the wiring of intrinsically safe
circuits must be observed.

The indicating and adjustment modules indicates faults via error codes
and text messages. The following table describes the error codes with
status according to NE 107 and gives information on the causes of
failure and their removal:

Status accor-
ding to NE 107

Error code

Text message

Cause/Rectification

Failure

E

013

no measured value available

Sensor in boot phase

no measured value available

Sensor does not find an echo, e.g.
due to faulty installation or wrong
parameter adjustment

no measured value available

Wrong sensor length entered

Error messages via the
indicating and adjust-
ment module

36

VEGAFLEX

63 • Profibus PA

8 Maintenance and fault rectification

31843

-EN

-100426