Scaling, Current output, Unit of measurement – VEGA VEGAFLEX 66 (-110…+400°C) 4 … 20 mA_HART - four-wire User Manual
Page 31: Language, Hart mode, Sensor length/sensor type, Gating out of false signals, The following values will be reset

l
S
caling
l
C
urrent output
l
U
nit of measurement
l
L
anguage
l
S
ensitivity
T
he following safety-relevant data are not read out or written:
l
HART
mode
l
PIN
l
SIL
l
S
ensor length/Sensor type
l
G
ating out of false signals
C
opy sensor data
C
opy sensor data?
Basic adjustment
I
f the function "Reset" is carried out, the sensor resets the
values of the following menu items to the reset values (see
chart):
1)
T
he following values will be reset:
Function
Reset value
M
ax. adjustment
D
istance, upper dead zone
M
in. adjustment - Rod/Coax version
D
istance, supplied sensor length
M
in. adjustment - Cable version
D
istance, lower dead zone
I
ntegration time ti
0
s
L
inearisation
linear
S
ensor-TAG
S
ensor
D
isplay
D
istance
C
urrent output - characteristics
4 … 20
mA
C
urrent output - max. current
20
mA
C
urrent output - min. current
4
mA
C
urrent output - failure
<
3.6
mA
A
pplication - rod/coax version
L
iquid
A
pplication - Cable version
S
olid
T
he values of the following menu items are not reset to the
reset values (see chart) with "Reset":
1)
S
ensor-specific basic adjustment.
Reset
VEGAFLEX
66 • 4 … 20
mA/HART - four-wire
31