beautypg.com

VEGA VEGAFLEX 66 (-110…+400°C) 4 … 20 mA_HART User Manual

Page 51

background image

Output variable

O

utput signal

4 … 20

mA/HART

HART

output values

-

HART

value (Primary Value)

D

istance to the level

-

HART

value (Secondary Value)

D

istance to the level - scaled

S

ignal resolution

1.6 µA

F

ailure signal current output (adjust-

able)

mA-value unchanged 20.5 mA, 22 mA,
<

3.6

mA (adjustable)

M

ax. output current

22

mA

L

oad

see load diagram under Power supply

D

amping (63 % of the input variable)

0 … 999

s, adjustable

F

ulfilled NAMUR recommendations

NE

43

Accuracy (similar to DIN EN 60770-1)

R

eference conditions according to DIN EN 61298-1

-

T

emperature

+

18 … +30 °C (+64 … +86 °F)

-

R

elative humidity

45 … 75 %

-

A

ir pressure

860 … 1060

mbar/86 … 106 kPa

(

12.5 … 15.4

psig)

Deviation in characteristics and characteristics

R

eference installation conditions

-

F

lange size

> DN

100

-

min. distance to installations

>

500

mm (19.69 in)

-

V

essel diameter

1

m (3.281 ft) centrical installaton

-

M

edium

W

ater

T

emperature drift (current output)

0.06 %/10 K

relating to the max. measuring

range

A

ccuracy

see diagrams

VEGAFLEX

66 • 4 … 20

mA/HART

51

10 S

upplement

32762

-

EN

-080716