3.2 Principle of operation

VEGASON

64 is an ultrasonic sensor for continuous level measure-

ment. It is particularly suitable for solids, but also good for liquids.

The transducer of the ultrasonic sensor transmits short ultrasonic
pulses to the measured product. These pulses are reflected by
product surface and received again by the transducer as echoes. The
running time of the ultrasonic pulses from emission to reception is
proportional to the distance and hence the level. The determined level
is converted into an appropriate output signal and outputted as
measured value.

Four-wire electronics with separate power supply.

Power supply of the signal output is via the Profibus DP/PA segment
coupler or VEGALOG 571 EP cards. A two-wire cable according to
Profibus specification serves as carrier of both power and digital data
signals for multiple sensors. The instrument profile of VEGASON 64
corresponds to profile specification version 3.0.

The GSD (instrument master files) and bitmap files necessary for
planning your Profibus-DP-(PA) communication network are available
from the download section on the VEGA homepage www.vega.com
under "Services - Downloads - Software - Pro

fibus". There you can

also find the appropriate certificates. In a PDM environment, an EDD
(Electronic Device Description) is also required to enable the full range
of sensor functions (also available as a download).A CD with the
appropriate files can be ordered via e-mail under [email protected] or
by phone from one of the VEGA agencies under the order number
"DRIVER

.S".

3.3 Operation

VEGASON

64 can be adjusted with different adjustment media:

l

with indicating and adjustment module

l

with the suitable VEGA DTM in conjunction with an adjustment
software according to the FDT/DTM standard, e.g. PACTware and
PC

l

with the adjustment program PDM

The entered parameters are generally saved in VEGASON 64,
optionally also in the indicating and adjustment module or in
PACTware.

Application area

Functional principle

Power supply and bus
communication

GSD/EDD

VEGASON

64 • Profibus PA