VEGA VEGAPULS 63 (≥ 2.0.0 - ≤ 3.8) Profibus PA User Manual
Page 49
-
flange DN 50, ANSI 2"
up to 10 m (32.81 ft)
-
Flange DN 80 … DN 150, Clamp 4",
ANSI
3" … 6"
up to 20 m (65.62 ft)
Reference conditions to measuring accuracy (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
-
Temperature
+
18 … +30 °C (+64 … +86 °F)
-
Relative humidity
45 … 75 %
-
Air pressure
860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)
Other reference conditions
-
Reflector
ideal reflector, e.g. metal plate 2 x 2 m
-
False reflections
Largest false echo 20 dB smaller than the useful
echo
Characteristics and performance data
Frequency
K-band (26 GHz technology)
Meas. interval approx.
1 s
Beam angle -3 dB
6)
-
Tri-Clamp 2"
18°
-
Clamp 3", 4"
10°
-
Bolting DN 50
18°
-
Bolting DN 80
10°
-
flange DN 50, ANSI 2"
18°
-
Flange DN 80 … DN 150, ANSI 3" … 6"
10°
Step response or adjustment time
7)
<
1 s (dependent on the parameter settings)
Max. level change
Adjustable up to 1 m/min. (dependent on the
parameter settings)
Max. emitted HF power of the antenna system
-
Pulse peak power
<
2 mW
-
Pulse duration
<
2 ns
-
Average power
<
5 µW
-
Average power with 1 m distance
<
200 nW/cm
2
Max. emitted HF power of the antenna system - Version with increased sensitivity
-
Pulse peak power
<
10 mW
-
Pulse duration
<
2 ns
-
Average power
<
25 µW
-
Average power with 1 m distance
<
1 µW/cm
2
6)
Corresponds to the range with 50 % of the emitted power
7)
Time to output the correct level (with max. 10 % deviation) after a sudden
level change.
VEGAPULS
63 • Profibus PA
49
10 Annex
28446
-EN
-120530