VEGA VEGASWING 61 - Relay (DPDT) User Manual
Page 28
28
9 Supplement
VEGASWING 61 • - Relay (DPDT)
29224-EN-130409
Ambient conditions
Ambient temperature on the housing
-40 … +70 °C (-40 … +158 °F)
Storage and transport temperature
-40 … +80 °C (-40 … +176 °F)
Process conditions
Measured variable
Limit level of liquids
Process pressure
-1 … 64 bar/-100 … 6400 kPa (-14.5 … 928 psig)
The process pressure is dependent on the process
fitting, for example Clam p or flange (see the following
diagrams)
Process pressure - Clamp fittings
Ʋ Clamp connection up to 2"
-1 … 16 bar/-100 … 1600 kPa (-14.5 … 232 psig)
max. 150 °C (302 °F)
Ʋ Clamp connection > 2"
-1 … 0.5 bar/-100 … 50 kPa (-14.5 … 7.25 psig)
> 150 °C (> 302 °F)
Maximum allowable operating pressure 100 bar/10000 kPa (1450 psig) or 1.5 times process
pressure
The function of the instrument is ensured up to an
operating pressure of 100 bar/10000 kPa (1450 psig)
at a maximum process temperature of +50 °C (+122 °F)
(only with threaded versions).
Process temperature (thread or flange
temperature)
Ʋ VEGASWING 61 of 316L/Hastelloy
C22 (2.4602)
-50 … +150 °C (-58 … +302 °F)
Process temperature (thread or flange
temperature) with temperature adapter
(option)
Ʋ VEGASWING 61 of 316L/Hastelloy
C22 (2.4602)
-50 … +250 °C (-58 … +482 °F)
Ʋ VEGASWING 61 enamelled
-50 … +200 °C (-58 … +392 °F)
Ʋ VEGASWING 61 with ECTFE coating -50 … +150 °C (-58 … +302 °F)
Ʋ VEGASWING 61 with PFA coating