VEGA VEGAPULS 62 (≥ 2.0.0 - ≤ 3.8) standpipe ver. 4 … 20 mA_HART two-wire User Manual
Page 49
S
pring-loaded terminals for wire cross-sec-
tion
>
2.5
mm² (AWG 14)
Electromechanical data - version IP 66/IP 68 (1 bar)
C
able entry
-
S
ingle chamber housing
1
x IP 68 cable gland M20 x 1.5; 1 x blind stopper
M
20
x 1.5
-
D
ouble chamber housing
1
x IP 68 cable gland M20 x 1.5; 1 x blind stopper
M
20
x 1.5; 1 x blind stopper M16 x 1.5
C
onnection cable
-
W
ire cross-section
0.5
mm² (AWG 20)
-
W
ire resistance
<
0.036 Ω/
m
-
T
ensile strength
<
1200 N (270
lbf)
-
S
tandard length
5
m (16.4 ft)
-
M
ax. length
1000
m (3280 ft)
-
M
in. bending radius
25
mm (0.984 in) with 25 °C (77 °F)
-
D
iameter approx.
8
mm (0.315 in)
-
C
olour - standard PE
B
lack
-
C
olour - standard PUR
B
lue
-
C
olour - Ex-version
B
lue
Indicating and adjustment module
V
oltage supply and data transmission
through the sensor
I
ndication
LC
display in dot matrix
A
djustment elements
4
keys
P
rotection
-
unassembled
IP
20
-
mounted into the sensor without cover
IP
40
M
aterials
-
H
ousing
ABS
-
I
nspection window
P
olyester foil
Power supply
O
perating voltage
-
N
on-Ex instrument
14 … 36 V DC
-
EE
x-ia instrument
14 … 30 V DC
-
EE
x-d-ia instrument
20 … 36 V DC
VEGAPULS
62
standpipe version • 4 … 20 mA/HART two-wire
49
10 S
upplement
28737
-EN
-090305