Version C is supplied in three parts:

l

Transducer

l

Transducer cable

l

electronics for wall mounting

Assemble the version as follows:

1

Loosen the screws on the fixing ring (4) with a hexagon spanner
(size 4), remove the transducer tube out of the swivelling holder

2

Mount the flange

3

Insert the transducer tube in the requested length from below into
the swivelling holder

4

Fasten with screws (4); torque max. 10 Nm

5

Remove the plug from below out of the connection head and plug
it into the socket of the transducer tube

6

Plug the connection head to the transducer tube. Do not squeeze
any cables. The cylinder is in the correct position, if the lower
notch (2) of the transducer tube is no longer visible

1

2

3

4

5

Fig. 6: Mounting - Version C
1

Groove for locking the connection head

2

Notch

3

Screw of swivelling holder (hexagon SW 13)

4

Grub screw, fixing ring (hexagon spanner - size 4)

5

Terminal screws, connection head (hexagon spanner - size 4)

7

Tighten the screws (5) on the cylinder moderately with an Allen
wrench (size 4)

8

Loosen terminal screw of the swivelling holder (3) with a fork
spanner (SW 13)

9

Direct the sensor with the swivelling holder to the measured
product

10 Fasten the swivelling holder with screw (3), torque max. 20 Nm

Mounting - Version C

VEGASON

65 • Profibus PA

15

4 Mounting

28788

-

EN