VEGA VEGASON 65 4 … 20 mA_HART four-wire User Manual
Page 54
Electromechanical data
C
able entry
-
D
ouble chamber housing
l
1
x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),
1
x blind stopper M20 x 1.5; plug M12 x 1 for
VEGADIS
61 (
optional)
or:
l
1
x closing cap ½ NPT, 1 x blind stopper
½ NPT
,
plug M12 x 1 for VEGADIS 61
(
optional)
or:
l
1
x plug (depending on the version), 1 x blind
stopper M20 x 1.5; plug M12 x 1 for VEGADIS
61 (
optional)
S
pring-loaded terminals for wire cross-sec-
tion up to
2.5
mm² (AWG 14)
T
ransducer cable
10)
-
L
ength
5 … 300
m (16.4 … 984.3 ft)
-
D
iameter
7.2 … 7.6
mm (0.283 … 0.299 in)
Indicating and adjustment module
V
oltage supply and data transmission
through the sensor
I
ndication
LC
display in dot matrix
A
djustment elements
4
keys
P
rotection
-
unassembled
IP
20
-
mounted into the sensor without cover
IP
40
M
aterials
-
H
ousing
ABS
-
I
nspection window
P
olyester foil
Power supply
O
perating voltage
-
N
on-Ex and Ex-d instrument
20 … 72 V DC, 20 … 253 V AC, 50/60 H
z
P
ower consumption
max. 4 VA; 2.1 W
Electrical protective measures
P
rotection
IP
66/IP 67
O
vervoltage category
III
P
rotection class
I
10)
W
ith separate version C and D.
54
VEGASON
65 • 4 … 20
mA/HART four-wire
10 S
upplement
28782
-EN
-081127