10 Supplement
VEGACAL 65 • 4 … 20 mA/HART - two-wire
30033-EN-130917
Ʋ PTFE partly insulated: ø 8 mm
(0.315 in)
10 KN (2248 lbf)
Max. torque (process fitting - thread)
100 Nm (73 lbf ft)
Output variable
Output signal
4 … 20 mA/HART
HART output values
Ʋ HART value (Primary Value)
Capacitance
Ʋ HART value (Secondary Value)
Capacitance - scaled
Resolution
1.6 µA
Failure signal current output (adjustable) mA value unchanged 20.5 mA, 22 mA, < 3.6 mA (adjust-
able)
Current limitation
22 mA
Load
see load diagram under Power supply
Damping (63 % of the input variable)
0 … 999 s, adjustable
Rise time
500 ms (ti: 0 s, 0 … 100 %)
Met NAMUR recommendation
NE 43
Input variable
Measured variable
level of non-conductive liquids and solids
Measuring principle
phase-selective admittance processing (PSA)
Measuring range
0 … 3000 pF
Measuring frequency
270 kHz
Accuracy (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
Ʋ Temperature
+18 … +30 °C (+64 … +86 °F)
Ʋ Relative humidity
45 … 75 %
Ʋ Air pressure
860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)
Temperature error
Ʋ < 120 pF
< 1 pF
Ʋ > 120 pF
1 % of the current measured value
Linearity error
< 0.25 % of the complete measuring range
Ambient conditions
Ambient, storage and transport tempera-
ture
-40 … +80 °C (-40 … +176 °F)
Process conditions
Process pressure
Ʋ Standard
-1 … +64 bar/-100 … 6400 kPa (-14.5 … 928 psig)
Ʋ with screening tube adapter (PN1)
0 … +1 bar/0 … 100 kPa (0 … 14.5 psig)
Process temperature (thread or flange temperature)