VEGA VEGACAL 66 Foundation Fieldbus User Manual
Page 38
38
10 Supplement
VEGACAL 66 • Foundation Fieldbus
30038-EN-130917
Ʋ Signal
digital output signal, Foundation Fieldbus protocol
Ʋ Physical layer
according to IEC 61158-2
Channel Numbers
Ʋ Channel 1
Primary value
Ʋ Channel 2
Secondary value 1
Ʋ Channel 3
Secondary value 2
Transmission rate
31.25 Kbit/s
Current value
10 mA, ±0.5 mA
Damping (63 % of the input variable)
0 … 999 s, adjustable
Input variable
Measured variable
Level of conductive liquids and solids
Measuring principle
phase-selective admittance processing (PSA)
Measuring range
0 … 3000 pF
Measuring frequency
270 kHz
Accuracy (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
Ʋ Temperature
+18 … +30 °C (+64 … +86 °F)
Ʋ Relative humidity
45 … 75 %
Ʋ Air pressure
860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)
Temperature error
Ʋ < 120 pF
< 1 pF
Ʋ > 120 pF
1 % of the current measured value
Linearity error
< 0.25 % of the complete measuring range
Ambient conditions
Ambient, storage and transport tempera-
ture
-40 … +80 °C (-40 … +176 °F)
Process conditions
Process pressure
-1 … 40 bar/-100 … 4000 kPa (-14.5 … 580 psig)
Process temperature VEGACAL 66 of
316L
-50 … +150 °C (-58 … +302 °F)