2 principle of operation – VEGA VEGACAL 62 Profibus PA User Manual
Page 8
8
3 Product description
VEGACAL 62 • Profibus PA
30025-EN-130917
2
1
11
13
12
14
15
10
9
3
6
4
5
7
8
Fig. 2: Layout of the type label (example)
1 Instrument type
2 Product code
3 Approvals
4 Process and ambient temperature, process pressure
5 Power supply and signal output, electronics
6 Protection rating
7 Probe length
8 Order number
9 Serial number of the instrument
10 Material, wetted parts
11 Symbol of the device protection class
12 Reminder to observe the instrument documentation
13 ID numbers, instrument documentation
14 Notified authority for CE marking
15 Approval directives
With the serial number, you can access the delivery data of the instru-
In addition to the type label outside, you can also find the serial num-
ber on the inside of the instrument.
With the serial number, you can access the delivery data of the instru-
In addition to the type label outside, you can also find the serial num-
ber on the inside of the instrument.
3.2 Principle of operation
VEGACAL 62 is a level sensor with partly insulated probe for continu-
ous level measurement.
The electronics functions according to the admittance principle
(phase-selective admittance processing).
It is designed for industrial use in all areas of process technology and
can be applied in all areas of industrial process measurement.
Partly insulated probes such as VEGACAL 62 are preferably used in
bulk solids.
However, the probe can also be used in non-conductive liquids such
as e.g. oil.
Probe, measured product and vessel wall form an electrical capacitor.
The capacitance is influenced by three main factors.
Area of application
Functional principle