beautypg.com

VEGA D86_D87 Hydrostatic pressure transmitters User Manual

Page 12

background image

12

Pressure transmitters D86/D87

Accuracy (similar to DIN 16 086, DIN V 19 259 - 1 and IEC 770)

Deviation

Reference conditions (acc. to IEC 770)
- temperature

15°C … 30°C

- humidity

45 % … 75 %

- air pressure

86 kPa … 106 kPa

Determination of characteristics

limit point adjustment acc. to DIN 16 086

Characteristics

linear

Deviation in characteristics

incl. hysteresis and repeatability

- Turn Down 1 : 1

< 0.25 % with accuracy class 0.25
< 0.1 % with accuracy class 0.1

- Turn Down up to 1 : 5

typ. < 0.3 % with accuracy class 0.25
typ. < 0.1 % with accuracy class 0.1

- Turn Down up to 1 : 10

typ. < 0.4 % with accuracy class 0.25
typ. < 0.2 % with accuracy class 0.1

Influence of the ambient temperature

Average temperature coefficient
of the zero signal

1)

- Turn Down 1 : 1

< 0.15 %/10 K with accuracy class 0.25
< 0.05 %/10 K with accuracy class 0.1

- Turn Down up to 1 : 5

typ. < 0.225 %/10 K with accuracy class 0.25
typ. < 0.075 %/10 K with accuracy class 0.1

- Turn Down up to 1 : 10

typ. < 0.3 %/10 K with accuracy class 0.25
typ. < 0.1 %/10 K with accuracy class 0.1

Long-term stability

Long-term stability of the zero signal

2)

< 0.25 % per 2 years

Other actuating variables

Calibration position

upright, diaphragm points downwards

Influence of the installation position
- D86

< 0,2 mbar

- D87

< 5 mbar

Vibration resistance

mechanical vibrations with 4 g and

5 … 100 Hz, tested acc. to the regulations of
German Lloyd GL-characteristics 2

Product description

1)

In compensated temperature range of 0°C … +80°C, reference temperature 20°C.

2)

Acc. to IEC 770, item 6.3.2 relating to the nominal range.