beautypg.com

VEGA VEGASON 51K…53K User Manual

Page 20

background image

20

VEGASON 51K … 53K

CE-conformity

VEGASON series 50 ultrasonic sensors meet the protective regulations of EMVG
(89/336/EWG) and NSR (73/23/EWG). The conformity has been judged acc. to the following
standards:
EMVG Emission

EN 50 081 - 1: 1993

Susceptibility

EN 50 082 - 2: 1995

NSR

EN 61 010 - 1: 1993

Outputs and processings

Display indication

Indication

- optional integrated in the sensor, scalable

analogue and digital meas. value indication

- optional external meas. value indication,

powered by the sensor and up to 25 m away
from the sensor (also the adjustment module
MINICOM can be inserted)

Signal output

Signal output
- two-wire technology

4 … 20 mA (load dependent on supply
voltage, see load diagram page 17)

- four-wire technology

4 … 20 mA (load max. 500

)

Resolution of the 20 mA-signal

0,025 % of operating range

Failure of the 20 mA-signal

< 0,025 % of operating range

Two-wire technology:
The analogue 4 … 20 mA-output signal (measuring signal) is transmitted together with the
power supply via a two-wire line.

Four-wire technology:
Separate power supply.
The analogue 4 … 20 mA-output signal (measuring signal) is looped by a line separated
from the supply voltage.

Technical data