VEGA D80 Hydrostatic pressure transmitters User Manual
Page 37
Pressure transmitter D80
37
Saved sensors can be transferred later on to
other sensors.
For example, if you have a system with sev-
eral of the same storage vessels and identi-
cal sensors, it is sufficient to configure one
sensor and then transfer the settings to the
other sensors.
In this menu window, the actual setting (data
base) with date and time of the last system
configuration is displayed in the yellow win-
dow field. If you click on the serial number of
the sensor from which you want to adopt the
adjustments (in the yellow field), these sensor
settings will be transferred to the currently
connected sensor with "
Restore to".
Setup
Meas. data information
• Click to "
Meas. data info".
All available sensor information is displayed.
With the button "
reset" you can delete the
pressure and temperature history saved in
the sensor.
Backup
In the menu window "
Backup" the sensor with
the serial number is displayed. You can save
the sensor on the PC individually or in groups
with all adjustments in a directory of your
choice. A small text message can be added
to each backup.