VEGA D96_D97 Profibus PA User Manual
Page 8
8
Pressure transmitter D96/D97 (Profibus PA)
1)
Relating to the nominal range.
2)
In compensated temperature range of 0°C … +80°C, reference temperature 20°C.
3)
Acc. to IEC 770, item 6.1.2 relating to the nominal range.
Product description
Accuracy (similar to DIN 16 086, DIN V 19 259 - 1 and IEC 770)
Deviation
Reference conditions (acc. to IEC 770)
- temperature
15°C … 35°C
- rel. humidity
45 % … 75 %
- air pressure
860 mbar … 1060 mbar/86 kPa … 106 kPa
Determination of characteristics
limit point adjustment acc. to DIN 16 086
Characteristics
linear
Deviation in characteristics
1)
- Turn Down 1 : 1
< 0.25 % with accuracy class 0.25
< 0.1 % with accuracy class 0.1
- Turn Down up to 1 : 5
typ. < 0.3 % with accuracy class 0.25
typ. < 0.1 % with accuracy class 0.1
- Turn Down up to 1 : 10
typ. < 0.4 % with accuracy class 0.25
typ. < 0.2 % with accuracy class 0.1
Hysteresis
1)
and repeatability
1)
- D96 (ceramic measuring cell)
< 0.02 %
- D97 (metal measuring cell)
< 0.05 %
Influence of the ambient temperature
Average temperature coefficient
of the zero signal
1) 2)
- Turn Down 1 : 1
< 0.15 %/10 K with accuracy class 0.25
< 0.05 %/10 K with accuracy class 0.1
- Turn Down up to 1 : 5
typ. < 0.225 %/10 K with accuracy class 0.25
typ. < 0.075 %/10 K with accuracy class 0.1
- Turn Down up to 1 : 10
typ. < 0.3 %/10 K with accuracy class 0.25
typ. < 0.1 %/10 K with accuracy class 0.1
Long-term stability
Long-term drift of the zero signal
3)
- D96 (ceramic measuring cell)
< 0.1 % per 2 years
- D97 (metal measuring cell)
< 0.1 % per 2 years
Other actuating variables
Calibration position
upright, diaphragm points downwards
Influence of the installation position
- D96 (ceramic measuring cell)
< 0.2 mbar
- D97 (metallic measuring cell)
< 5.0 mbar
Vibration resistance
mechanical vibrations with 4 g and
5 … 100 Hz, tested acc. to the regulations of
German Lloyd GL-characteristics 2