2 enquiry as range – VEGA VEGASCAN 850 VEGA-ASCII signal output User Manual
Page 7
VEGA-ASCII signal output VEGASCAN 850
7
Enquiry of measured values
3.2 Enquiry as range
Low resolution
This telegram reads a DCS value out of VEGASCAN 850.
Answer from VEGASCAN 850
Start character: S1 = 1 character "=“
Start character: S2 = 1 character "#“
Total length of the telegram: 12 or 13 characters
Telegram from processing system to VEGASCAN 850
n (n= stated number) repetitions
1 2 3 4 5 6
3 positions in front of the comma
1 decimal point
1 position after the comma
"-“ with negative values
" " with positive values
valid measured value
1 2 3 4 5
‘FAULT’
Fault
Identifier: 1 character, "%“
T1 : Separator character = 1 character "L“
End identification: 1 character Chr (13) = CR = Carriage Return
Total length of the telegram: 9 characters
DCS
DCS
End
S1
no.
S2
value2
identification
Number of bytes
1
3
1
5-6
2
Range (ASCII)
=
001..060
#
PLS-Wert
oder Fault
%CR
Example
=
001
#
-067.3
%CR
DCS
End
Identifier
number
T1
Number
identification
Number of bytes
1
3
1
3
1
Range (ASCII)
%
001..060
L
001..060
CR
Example
%
001
L
001
CR
- VEGABAR 25 (24 pages)
- VEGAFLEX 55K (32 pages)
- VEGABAR 17 (36 pages)
- D80 Hydrostatic pressure transmitters (40 pages)
- VEGAMET 513 (52 pages)
- D81 Hydrostatic pressure transmitter (44 pages)
- VEGASTAB 690 (12 pages)
- VEGASWING 83 fail safe and signal conditioning instrument VEGATOR 526 Ex (28 pages)
- VEGACONNECT 2 (4 pages)
- D81 METEC Hydrostatic pressure transmitter (44 pages)
- VEGAPULS 45 4 … 20 mA; HART compact sensor (72 pages)
- VEGAMET 515 (68 pages)
- VEGAFLEX 52K (32 pages)
- Capacitive electrodes EK Profibus PA (56 pages)
- VEGATOR 631 S1 (4 pages)
- VEGAPULS 63 (≥ 2.0.0 - ≤ 3.8) 4 … 20 mA_HART - two-wire (68 pages)
- Capacitive electrodes EK 4 … 20 mA - Compact (40 pages)
- VEGABAR 65 4 … 20 mA_HART (76 pages)
- VEGABAR 64 Profibus PA (80 pages)
- VEGABAR 44 4 … 20 mA (28 pages)
- VEGADIS 50 Ex (8 pages)
- VEGAPULS 51V…54V (96 pages)
- VEGAFLEX 54P (Profibus PA) (48 pages)
- VEGASEL 643 (16 pages)
- VEGAPULS 56K (68 pages)
- VEGAFLEX 54K (36 pages)
- VEGABAR 41 HART (44 pages)
- VEGATOR 537 Ex (16 pages)
- VEGACAP 82-98 (36 pages)
- VEGADIS 371 Ex (20 pages)
- Capacitive electrodes EL Profibus PA (56 pages)
- VEGAFLEX 52V (36 pages)
- VEGAFLEX 51K (32 pages)
- VEGALOG 571 EP (Ex) (32 pages)
- D84 Hydrostatic pressure transmitter (44 pages)
- VEGABAR 64 Foundation Fieldbus (80 pages)
- VEGABAR 61 4 … 20 mA_HART (80 pages)
- VEGASCAN 850 (92 pages)
- VEGASON 54P…56P Profibus PA (68 pages)
- D77 Hydrostatic pressure transmitters (24 pages)
- VEGATRENN 546 Safety barrier type 146 (8 pages)
- VEGASON 71 - 1 … 75 - 1 (52 pages)
- VEGABAR 40 HART (40 pages)
- Capacitive electrodes EL … (44 pages)
- VEGABAR 24 (28 pages)