3 technical data – VEGA VEGABAR 24 User Manual
Page 5
VEGABAR 24
5
Product description
1.3 Technical data
Mechanical data
Materials, wetted parts
Process connection
stainless steel 1.4571, PVDF
1)
Diaphragm
ceramic (99,5 % Oxydeceramic)
Seal measuring cell
Viton, Kalrez, EPDM
Materials, non-wetted parts
Housing
high resistance plastic PBT (Polyester)
Earth terminal
stainless steel 1.4305
Indicating module window
glass
Weight
VEGABAR
approx. 0,8 … 8 kg (dependent on process
connection)
Adjustment and indicating elements
Adjustment of the basic functions
2 keys, 1 rotating switch
Menu-guided adjustment with
additional function
- adjustment elements
4 key
- indicating elements
DOT-matrix display, 3 lines with 7 figures each
Indicating module
LC-display with
- bar graph (20 segments)
- digital value (4-digit)
- tendency indicators for raising or lowering
values
Electrical data
Connection
Cable entry
PG 13,5 (for cable-ø 5 … 10 mm)
Screw terminals
for conductor cross-sections of up to
2,5 mm
2
Protective measures
Protection
1)
IP 65
Protection class
III
Overvoltage category
III
CE-conformity
VEGABAR 24 meets the requirements of EMC (89/336/EWG) and NSR (73/23/EWG). The
conformity has been judged acc. to the following standards
EMC
Emission
EN 50 081 - 1: 1992
Susceptibility
EN 50 082 - 2: 1995
NSR
EN 61 010 - 1: 1993
1)
Use of a suitable seal for the cable in the PG is necessary for maintaining the housing protection. If the seal
used does not fit the cable, it should be replaced by a suitable one.