beautypg.com

VEGA VEGABAR 24 User Manual

Page 18

background image

18

VEGABAR 24

Set-up

Adjustment taking the current pres-
sure into account (live adjustment)

Live adjustment comprises two steps:
1 Adjustment of zero
2 Adjustment of span

The appropriate output current is displayed
via the DOT-matrix.

1 Adjustment of zero
(e.g. process pressure zero or vessel
empty)

Set the current value to 4,000 mA
with the “+“ or “–“-key. Then push
the “OK“-key.

2 Adjustment of span
(e.g. process pressure or vessel level max.)

Set the current value to 20,000 mA
with the “+“ or “–“-key. Then push
the “OK“-key.

Note:
- A modification of zero does not influence

the adjusted span.

- “+“ and “–“ are pushed simultaneously for

standard adjustment of zero/span. The
value jumps directly to 4 mA/20 mA.

- In the case of a high turn down the “+“ and

“-“-key should generally be pressed simul-
taneously due to the resolution.

- When pushing the “+“ or “–“-keys individu-

ally, the output current remains at the last
value, it takes on the adjusted value only
after storing with the “OK“-key.

- It is also possible to adjust currents for

partial fillings or partial pressures, e.g.
8 mA for 25 % and 16 mA for 75 %. Then
VEGABAR automatically calculates the
current values for 0 % or 100 % (only pos-
sible with >3,3 %).

Adjustment without taking the current
pressure into account (dry adjustment)

Adjustment without pressure comprises four
steps:
1 Selection of the unit in which the adjustment

is to be carried out

2 Offset correction
3 Adjustment of zero
4 Adjustment of span

The offset correction (only with overpressure)
defines the reference basis for the measure-
ment. It can be carried out:
- before or after the adjustment of zero and

span

- before or after installation of VEGABAR.

VEGABAR must be unpressurised when the
offset correction is carried out!

Adjustment without pressure can be carried
out assembled or unassembled (e.g. work-
room). A currently existing pressure has no
significance for the adjustment.

1 Enquiry of the unit

Choose the unit with the “+“ or “–“.
Save the selected unit with the
“OK“-key.

2 Adjustment of zero

Adjust zero with the “+“ or “–“-key
and save with the “OK“-key.

4 Adjustment of span

Adjust span with the “+“ or “–“-key
and save with the “OK“-key.

Zero

4,000

mA

Span

20,000

mA

Zero

0,0

mbar

Span

100,0

mbar

Adjust

without

press.

Unit

0,0

mbar

psi
kPa