beautypg.com

2 connection procedure – VEGA VEGABAR 75 4 … 20 mA_HART User Manual

Page 16

background image

5

.2 Connection procedure

P

roceed as follows:

1

W

ire the connection cable up to the connection compart-

ment. The bending radius must be at least 25 mm.

2)

2

C

onnect the wire ends to the screw terminals according to

the wiring plan

P

roceed as follows:

1

S

nap connection housing onto the carrier rail or screw it to

the mounting plate

2

L

oosen the cover screws and remove the cover

3

I

nsert the cable through the cable entry into the connection

housing housing

4

L

oosen the screws with a screwdriver

5

I

nsert the wire ends into the open terminals according to

the wiring plan

6

T

ighten the screws with a screwdriver

7

C

heck the hold of the wires in the terminals by lightly

pulling on them

8

T

ighten the compression nut of the cable entry. The seal

ring must completely encircle the cable

9

C

onnect the supply cable according to steps 3 to 8

10 S

crew the housing cover back on

T

he electrical connection is finished.

2)

T

he connection cable is already preconfectioned. After shortening the ca-

ble, fasten the type plate with support again to the cable.

Direct connection

Via VEGABOX 01 or VEGADIS
12

16

VEGABAR

75 - 4 … 20

mA/HART

C

onnecting to power supply

28433

-EN

-070718