Pressure transmitter D90 (Profibus PA)

7

Product description

1)

The use of a properly fitting cable seal in the cable connection is necessary for maintaining the housing
protection. If the seal used does not fit the cable, it should be replaced by a suitable one.

Electrical data

Adjustment ranges

Zero

adjustable from -20 … +95 % of nominal range

Span

adjustable from 3.3 … 120 % of nominal range

Supply and signal circuit

Supply voltage

9 … 32 V DC
(delivered by the segment coupler)

Current consumption

10 mA ± 1 mA

Integration time

0 … 50 s

Rise time

70 ms (ti = 0 sec; 0 – 63 %)

Connection cable, processing unit – Profibus PA

Cable entry

M20 x 1.5 (for cable ø 5 … 9 mm)

Screw terminals

for conductor cross-sections of up to 2.5 mm

2

Recommended bus cable types

- SINEC 6XV 830-5AH10 (Siemens AG)

- SINEC L2 6XV1 830-35H10 (Siemens AG)
- 3079 A (Belden)

Connection cable, processing unit – Transmitter

Cable entry

M20 x 1.5 (for cable ø 5 … 9 mm)

Screw terminals

for conductor cross-sections of up to 2.5 mm

2

Transmitter
- with fixed cable outlet

3-wire screened with breather capillaries

- with cable connection terminals

3-wire screened

Protective measures

1)

Protection
- transmitter with fixed cable outlet

IP 68

- transmitter with terminal housing

IP 66/IP 67

- transmitter with plug connection

IP 66/IP 67

- processing unit

IP 66/IP 67

Protection class

III

Overvoltage category